Brain Cancer Research - Symptoms, Benign and Malignant Tumors, Gliomas, Treatment

Brain Cancer Research Today is a free monthly online journal that collates and summarizes the latest research about Brain Cancer, including details on symptoms, benign and malignant tumors, gliomas, treatment.


Brain Cancer Research Today

Home

View Latest Issue

Information About Brain Cancer

Books on Brain Cancer

Advertising in Research Today

View Other Research Today Publications



Optimization of megavoltage CT scan registration settings for brain cancer treatments on tomotherapy.

Woodford C, Yartsev S, Van Dyk J

London Regional Cancer Program, London Health Sciences Centre, 790 Commissioners Road East, London, Ontario, Canada.

This study aims to determine the settings that provide the optimal clinical accuracy and consistency for the registration of megavoltage CT (MVCT) with planning kilovoltage CT image sets on the Hi-ART tomotherapy system. The systematic offset between the MVCT and the planning kVCT was determined by registration of multiple MVCT scans of a head phantom aligned with the planning isocentre. Residual error vector lengths and components were used to quantify the alignment quality for the phantom shifted by 5 mm in different directions obtained by all 27 possible combinations of MVCT inter-slice spacing, registration techniques and resolution. MVCT scans with normal slices are superior to coarse slices for registration of shifts in the superior-inferior, lateral and anterior-posterior directions. Decreasing the scan length has no detrimental effect on registration accuracy as long as the scan lengths are larger than 24 mm. In the case of bone technique and fine resolution, normal and fine MVCT scan slice spacing options give similar accuracy, so normal mode is preferable due to shorter procedure and less delivered dose required for patient set-up. A superior-inferior field length of 24-30 mm, normal slice spacing, bone technique, and fine resolution is the optimum set of registration settings for MVCT scans of a Rando head phantom acquired with the Hi-ART tomotherapy system, provided the registration shifts are less than 5 mm.

Published 3 April 2007 in Phys Med Biol, 52(8): N185-93.
Full-text of this article is available online (may require subscription).

Place a permanent text-link or advertisement here for just US$15.

© 2004-2008 Brain Cancer Research Today. All Rights Reserved.



Brain Cancer Research Today Archive:

Volume 1 (2004)
  Issue 1 (August)
  Issue 2 (September)
  Issue 3 (October)
  Issue 4 (November)
  Issue 5 (December)

Volume 2 (2005)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 3 (2006)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 4 (2007)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 5 (2008)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)



Brain Cancer Books

ONE YEAR TO LIVE? A Nobody's Guide to Surviving Cancer

ONE YEAR TO LIVE? A Nobody's Guide to Surviving Cancer